Balancing brightness and bleed-through
A wider emission band can improve signal, but it may also collect unwanted fluorescence from nearby channels. Start from the fluorophore spectrum and detector sensitivity.
OPTOStokes is the optical filter brand at Optofilters.com, supplying stock and custom-coated filters for UV, VIS, NIR and IR applications.
Review matched fluorescence filter sets for common microscopy fluorophores. The tables help compare excitation bands, emission bands, dichroic mirrors, and practical replacement options for imaging systems.
Additional normalized fluorescence microscopy options. Confirm the fluorophore path, blocking curve, substrate, and mechanical format before production.
| Model | Application | CWL (nm) | FWHM (nm) | Transmittance | Blocking / Optical Density | Transmission Range (nm) |
|---|---|---|---|---|---|---|
| STOS-LP515 | Fluorescence microscopy / FITC | Not specified | Not specified | T≥92% | ODavg > 6 at 200-505 nm | 525-800 |
| STOS-450NF | Fluorescence microscopy | 450 ± 8 | 50 ± 5 | T≥90% | OD ≥ 5 at 445-455 nm | 405-420&480-700 |
| STOS-460NF | Fluorescence microscopy | 460 ± 10 | 65 ± 5 | T≥90% | OD ≥ 3 at 440-480 nm | 405-425&500-700 |
| STOS-470NF | Fluorescence microscopy | 470 ± 8 | 45 ± 5 | Tave≥95% | T < 5% at 465-475 nm | 350-425&505-700 |
| STOS-485NF | Fluorescence microscopy | 485 ± 20 | 190 ± 10 | T≥90% | OD ≥ 4 at 420-550 nm | 350-380&600-1100 |
| STOS-528NF | Fluorescence microscopy | 528 ± 5 | 24 ± 5 | T≥90% | OD ≥ 2 at 527-529 nm | 400-505&560-700 |
| STOS-530NF | Fluorescence microscopy | 530 ± 10 | 90 ± 10 | Tave≥95% | OD ≥ 5 at 500-560 nm | 400-470&590-800 |
| STOS-560NF | Fluorescence microscopy | 560 ± 10 | 105 ± 10 | Tave≥95% | OD ≥ 2 at 535-585 nm | 350-480&650-750 |
| STOS-575NF | Fluorescence microscopy | 575 ± 5 | 35 ± 5 | Tave≥95% | OD ≥ 2 at 570-580 nm | 430-550&600-800 |
| STOS-590NF | Fluorescence microscopy | 590 ± 5 | 35 ± 5 | T≥90% | OD ≥ 2 at 585-595 nm | 430-560&620-800 |
| STOS-600NF | Fluorescence microscopy | 600 ± 10 | 80 ± 5 | Tave≥90% | OD ≥ 2 at 630-670 nm | 380-595&725-1100 |
| STOS-625NF | Fluorescence microscopy | 625 ± 10 | 70 ± 5 | Tave≥95% | OD ≥ 6 at 610-640 nm | 500-580&675-700 |
| STOS-730NF | Fluorescence microscopy | 730 ± 15 | 150 ± 10 | Tave≥90% | OD ≥ 6 at 675-790 nm | 420-645&825-1100 |
| STOS-765NF | Fluorescence microscopy | 765 ± 15 | 170 ± 10 | T≥90% | OD ≥ 5 at 780-820 nm | 450-665&880-1100 |
| STOS-770NF | Fluorescence microscopy | 770 ± 15 | 145 ± 10 | T≥85% | OD ≥ 5 at 710-810 nm | 450-680&870-1100 |
| STOS-775NF | Fluorescence microscopy | 775 ± 15 | 130 ± 10 | Tave≥90% | OD ≥ 3 at 730-810 nm | 415-690&850-900 |
| STOS-810NF | Fluorescence microscopy | 810 ± 20 | 200 ± 10 | T≥90% | OD ≥ 2 at 730-870 nm | 380-670&930-1100 |
| STOS-1080NF | Fluorescence microscopy | 1080 ± 10 | 60 ± 5 | Tave≥95% | OD ≥ 2 at 1075-1085 nm | 900-1040&1120-1700 |
Choose the fluorophore first, then check whether the excitation and emission passbands fit the dye spectrum and light source. Confirm dichroic transition, microscope holder size, filter thickness, and whether neighboring channels need tighter blocking.
OPTOStokes fluorescence microscope filters support common fluorophores used in widefield and confocal microscopy, including DAPI, FITC/EGFP, TRITC/PE, and CY5. This page lists individual excitation and emission filters by channel; a complete microscope configuration can add a matched dichroic mirror selected separately for the instrument and optical path.
Choose the fluorophore first, then review the matched excitation and emission filters. Shared models are listed once in the product table and may serve more than one channel.
| Fluorophore | Excitation Model | Excitation CWL / FWHM | Emission Model | Emission CWL / FWHM |
|---|---|---|---|---|
| DAPI | STO-375-28 | 375 ± 3 nm / 28 ± 3 nm | STO-460-55 | 460 ± 3 nm / 55 ± 3 nm |
| CFP | STO-430-24 | 430 ± 3 nm / 24 ± 3 nm | STO-480-30 | 480 ± 3 nm / 30 ± 3 nm |
| FITC | STO-470-40 | 470 ± 3 nm / 40 ± 3 nm | STO-535-40 | 535 ± 3 nm / 40 ± 3 nm |
| GFP | STO-480-30 | 480 ± 3 nm / 30 ± 3 nm | STO-535-40 | 535 ± 3 nm / 40 ± 3 nm |
| YFP | STO-485-20 | 485 ± 3 nm / 20 ± 3 nm | STO-530-20 | 530 ± 3 nm / 20 ± 3 nm |
| CY3/TRITC | STO-530-40 | 530 ± 3 nm / 40 ± 3 nm | STO-605-55 | 605 ± 3 nm / 55 ± 3 nm |
| mCherry | STO-560-40 | 560 ± 3 nm / 40 ± 3 nm | STO-635-60 | 635 ± 3 nm / 60 ± 3 nm |
| CY5 | STO-624-40 | 624 ± 3 nm / 40 ± 3 nm | STO-694-40 | 694 ± 3 nm / 40 ± 3 nm |
| CY5.5 | STO-665-30 | 665 ± 3 nm / 30 ± 3 nm | STO-725-40 | 725 ± 3 nm / 40 ± 3 nm |
| CY7 | STO-715-70 | 715 ± 3 nm / 70 ± 3 nm | STO-800-60 | 800 ± 3 nm / 60 ± 3 nm |
The latest reviewed source contains ten application channels and eighteen unique bandpass models. Public values use the confirmed microscopy transmission specification of T > 91%.
| Model | Role / Source Channel | Fluorophore Channels | CWL (nm) | FWHM (nm) | Transmittance (T) | Blocking | Standard Formats | Maximum Size (mm) |
|---|---|---|---|---|---|---|---|---|
| STO-375-28 | Excitation — DAPI | DAPI | 375 ± 3 | 28 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-430-24 | Excitation — CFP | CFP | 430 ± 3 | 24 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-460-55 | Emission — DAPI | DAPI | 460 ± 3 | 55 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-470-40 | Excitation — FITC | FITC | 470 ± 3 | 40 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-480-30 | Excitation — GFP; Emission — CFP | GFP, CFP | 480 ± 3 | 30 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-485-20 | Excitation — YFP | YFP | 485 ± 3 | 20 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-530-20 | Emission — YFP | YFP | 530 ± 3 | 20 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-530-40 | Excitation — CY3/TRITC | CY3/TRITC | 530 ± 3 | 40 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-535-40 | Emission — FITC; Emission — GFP | FITC, GFP | 535 ± 3 | 40 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-560-40 | Excitation — mCherry | mCherry | 560 ± 3 | 40 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-605-55 | Emission — CY3/TRITC | CY3/TRITC | 605 ± 3 | 55 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-624-40 | Excitation — CY5 | CY5 | 624 ± 3 | 40 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-635-60 | Emission — mCherry | mCherry | 635 ± 3 | 60 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-665-30 | Excitation — CY5.5 | CY5.5 | 665 ± 3 | 30 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-694-40 | Emission — CY5 | CY5 | 694 ± 3 | 40 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-715-70 | Excitation — CY7 | CY7 | 715 ± 3 | 70 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-725-40 | Emission — CY5.5 | CY5.5 | 725 ± 3 | 40 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 | |
| STO-800-60 | Emission — CY7 | CY7 | 800 ± 3 | 60 ± 3 | T>91% | ODavg > 6 at 350-800 nm (out-of-band; passband excluded where applicable); OD > 3 at 800-1050 nm (out-of-band; passband excluded where applicable) | 57 × 57 |
Mechanical variants are listed once here and shared by the microscope filter models above. Review the optical performance first, then select the format that matches your instrument.
| Format Family | Standard Format | Shape | Mounting | Clear Aperture | Retainer Tolerance | Size Tolerance | Substrate | Applies To |
|---|---|---|---|---|---|---|---|---|
| 1 standard format · Round · mounted · 1.0 mm substrate | Ø25 × 3.5 mm, mounted | Round | mounted | Ø22.0 mm | +0.0/-0.1 mm | Not specified | 1.0 mm (±0.1 mm) | 18 fluorescence microscope models |
Typical configuration: Ø25 × 3.5 mm mounted filter, Ø22.0 mm clear aperture, 1.0 mm substrate, and 57 × 57 mm maximum available size where specified by the source.
A wider emission band can improve signal, but it may also collect unwanted fluorescence from nearby channels. Start from the fluorophore spectrum and detector sensitivity.
Common microscope filter sets should be checked against the light source, camera response, and dichroic transition rather than dye name alone.
A dichroic transition that is too close to the emission band can reduce useful signal or increase channel leakage in multi-color imaging.
This catalog page lists individual excitation and emission filters by fluorophore. A complete microscope configuration may add a matched dichroic mirror selected separately for the instrument and optical path.
Often yes, if diameter, thickness, and optical orientation match the holder. Confirm mechanical dimensions before ordering.
The page covers common channels such as DAPI, FITC, EGFP, TRITC, PE, CY5, and related microscopy dyes.
Quality & Compliance
For OEM programs, laboratory instruments, and industrial systems, OPTOStokes keeps quality, material compliance, and engineering support visible from the start of supplier evaluation.
Process control, inspection records, and repeatable coating workflows for production optical components.
Compliance documentation can be prepared for qualified projects and regulated supply chains.
Wavelength, FWHM, substrate, AOI, size, and mounting requirements can be reviewed with engineering.
Share target wavelength, blocking level, transmittance, and drawing requirements for a focused quote.
Our engineers will help you select the perfect optical solution.