FITC fluorescence filter set: BP470/40 excitation+DM505 spectroscopy+BP535/40 emission

FITC fluorescence filter set: BP470/40 excitation+DM505 spectroscopy+BP535/40 emission

The FITC fluorescence filter set consists of BP470/40 excitation plate, DM505 dichroic mirror, and BP535/40 emission plate, with a blocking coverage of 200-1040nm. The size and mounting ring can be customized. Please feel free to inquire about the optical path parameters.
Products Description

FITC fluorescence filter set: BP470/40 excitation+DM505 spectroscopy+BP535/40 emission

FITC fluorescence filter set: BP470/40 excitation+DM505 spectroscopy+BP535/40 emission(pic1)

This is a set of three piece combinations for fluorescence microscopy FITC channels: BP470/40 limits the excitation light, DM505 reflects the excitation light to the sample at 45 ° and allows the emission light to pass through, and BP535/40 performs the final emission band selection in front of the detector. The fluorescence channel cannot be produced by a single filter, and selecting three filters together is the normal practice for such applications.

Technical specifications

BP470/40 (excitation)BP535/40 (emission)

FITC fluorescence filter group three piece specification comparison
ParametersBP470/40(EX)DM505 (dichroic mirror)BP535/40(EM)
Passband and transmittance455–485nm,Tave> 90%515–700nm,T> 95%515–550nm,Tave> 90%
Center wavelength470±2nm—535±2nm
Bandwidth FWHM40±2nm—40±2nm
Blocking and Reflecting Bands200–420nm、500–750nm:ODave> 6(OD>5); 750–1040nm:OD> 3430–490nm,T< 2%350–505nm、565–800nm:ODave> 6(OD>5); 800–1040nm:OD> 3
Incident angle0°45°0°

Why do FITC channels need to be matched with three pieces

The excitation peak of FITC is about 495nm, and the emission peak is about 519nm, separated by only about twenty nanometers. No single filter can separate the excitation light and emission light. The division of labor of the filter block is: the excitation plate first limits the incident short wave, the dichroic mirror divides the optical path into reflection and transmission according to wavelength, and the emission plate then blocks the residual excitation light and sample spontaneous background outside the detector.

The 45 ° marked on DM505 is not an installation habit, but a design angle. Its cutoff position only holds at this angle, and when the assembly is biased a few degrees, the separation degree around 490nm will change - this is most easily overlooked when modifying the frame or mixing filter blocks.

Selection Example: A set of FITC observation optical path with conventional objective lens, the requirement is to excite FITC and collect emission near 519nm.. Select BP470/40 to cover the excitation band main body, BP535/40 to cover the emission peak, and DM505 to complete the separation at 45 °. There are three boundary conditions: all three are defined as 0 °/45 ° collimated light, which is not applicable when there is a significant convergence angle in the optical path; The actual installation angle deviation of the dichroic mirror by 45 ° will move the cutoff edge; The BP470/40 passband ends at 485nm, located on the short wave side of the FITC excitation peak at 495nm. If the excitation efficiency is insufficient, it is necessary to evaluate whether to adjust the center wavelength of the excitation plate upwards. When verifying, do not only measure a single piece, but also measure the excitation and emission spectra of the entire group after installing a filter block.

The easiest pit to step on: the spectral boundaries of the three pieces have almost no margin

Calculate according to the nominal value: the long wave edge of BP470/40 reaches 485nm, the reflection band of DM505 ends at 490nm, leaving only 5nm in the middle; the transmission band of DM505 starts from 515nm, and the short wave edge of BP535/40 is exactly 515nm, with both ends directly connected. This is not a parameter error, it is a deliberate tight fit made by the fluorescence filter block to suppress cross color distortion.

The problem is that the margin is small, and any deviation in the loop will be exposed: the dichroic mirror is misaligned, the excitation plate is not working at 0 °, and the batch tolerance has reached the edge. As a result, the excitation light leaks into the emission channel, the image background is raised, and weak signals are submerged. However, each individual retest is qualified. The correct approach is to conduct a whole group acceptance - compare the signal-to-noise ratio before and after mounting with the same fluorescent sample, and confirm the actual installation angle of the dichroic mirror separately, rather than submitting a single spectral report.

Customization and matching

All three pieces can be processed into circular, square, or irregular shapes, equipped with snap rings, or delivered as a whole filter block assembly; It is recommended to confirm the specific dimensions, thickness, and collar structure together during the inquiry to avoid interference with the frame after the lens is in place. Another reminder: BP470/40 and BP535/40 have OD>in the 800-1040nm range; 3, not OD6. If the detector still responds in near-infrared, the suppression intensity in this section needs to be clearly stated during selection and evaluated separately if necessary.

To confirm if this combination matches your fluorophore and optical path, you can send the fluorophore name, excitation and emission bands, filter block model or installation size, and estimated quantity to sales@optofilters.com 。 If you want to see the configuration of other channels first, you can refer to<a href=“ https://www.optofilters.com/product/FluorescenceFilterSets.html Fluorescent filter group; when replacing one of them separately,<a ref= https://www.optofilters.com/product/BandpassFilters.html Bandpass filters can be classified by center wavelength and bandwidth.